Please use this identifier to cite or link to this item: http://elar.urfu.ru/handle/10995/130819
Title: Portable Device for Potentiometric Determination of Antioxidant Capacity
Authors: Ivanova, A. V.
Markina, M. G.
Issue Date: 2023
Publisher: Multidisciplinary Digital Publishing Institute (MDPI)
Citation: Ivanova, A & Markina, M 2023, 'Portable Device for Potentiometric Determination of Antioxidant Capacity', Sensors, Том. 23, № 18, 7845. https://doi.org/10.3390/s23187845
Ivanova, A., & Markina, M. (2023). Portable Device for Potentiometric Determination of Antioxidant Capacity. Sensors, 23(18), [7845]. https://doi.org/10.3390/s23187845
Abstract: For the first time, a prototype of a portable device for the potentiometric determination of antioxidant capacity based on a new measurement principle is proposed. A feature of the approach is the use of an electrochemical microcell with separated spaces and two identical electrodes with immobilized reagents. An antioxidant solution is introduced into one half-cell, and the antioxidants interact with the reagents. The other half-cell contains only reagents. The potential difference between the electrodes is due to the change in the ratio of the oxidized and reduced form of the reagents, which occurs as a result of the reaction with the antioxidants in one of the half-cells and is related to their concentration. The range of linearity of the microcell with immobilized reagents is 40–4000 μM-eq, and the limit of detection is 20 μM-eq. The device was successfully tested in the analysis of standard antioxidant solutions. The recoveries were (92–113)%, and the relative standard deviation did not exceed 15%. A good correlation was found between the data obtained by the approach and the potentiometric method in a macrocell for fruit juice analysis. Pearson’s coefficient for the obtained experimental data was 0.9955. The proposed portable device is promising and can be used in field conditions. © 2023 by the authors.
Keywords: ANTIOXIDANT CAPACITY
MICROCELL
PORTABLE DEVICE
POTENTIOMETRY
SCREEN-PRINTED ELECTRODE
CHEMICAL REACTIONS
ELECTROCHEMICAL ELECTRODES
FRUIT JUICES
POTENTIOMETERS (ELECTRIC MEASURING INSTRUMENTS)
ANTIOXIDANT CAPACITY
ELECTROCHEMICAL MICRO-CELL
HALF CELLS
IMMOBILIZED REAGENTS
MICRO CELL
PORTABLE DEVICE
POTENTIAL DIFFERENCE
POTENTIOMETRICS
POTENTIOMETRY
SCREEN PRINTED ELECTRODES
ANTIOXIDANTS
ANTIOXIDANT
PROCEDURES
VOLTAMMETRY
ANTIOXIDANTS
ELECTRODES
POTENTIOMETRY
URI: http://elar.urfu.ru/handle/10995/130819
Access: info:eu-repo/semantics/openAccess
cc-by
License text: https://creativecommons.org/licenses/by/4.0/
SCOPUS ID: 85172739825
WOS ID: 001075240800001
PURE ID: 45662308
ISSN: 1424-8220
DOI: 10.3390/s23187845
Sponsorship: Russian Science Foundation, RSF: 20-13-00142
This research was funded by a grant from the Russian Science Foundation No. 20-13-00142, https://rscf.ru/project/20-13-00142/ (accessed on 6 June 2023).
RSCF project card: 20-13-00142