Please use this identifier to cite or link to this item: http://elar.urfu.ru/handle/10995/117966
Title: Calculation of NaI(Tl) detector efficiency using 226Ra, 232Th, and 40K radioisotopes: Three-phase Monte Carlo simulation study
Authors: Tekin, H. O.
Almisned, G.
Issa, S. A. M.
Zakaly, H. M. H.
Kilic, G.
Ene, A.
Issue Date: 2022
Publisher: De Gruyter Open Ltd
Citation: Calculation of NaI(Tl) detector efficiency using 226Ra, 232Th, and 40K radioisotopes: Three-phase Monte Carlo simulation study / H. O. Tekin, G. Almisned, S. A. M. Issa et al. // Open Chemistry. — 2022. — Vol. 20. — Iss. 1. — P. 541-549.
Abstract: Thallium-activated sodium iodide (NaI(Tl)) detectors can be used in gamma cameras, environmental radiation assessments, including radiation emission levels from nuclear reactors, and radiation analysis equipment. This three-phase investigation aimed to model a standard NaI(Tl) detector using the Monte Carlo N-Particle eXtended (MCNPX) general-purpose Monte Carlo simulation techniques. Accordingly, a standard NaI(Tl) detector was designed along with the required properties. Next a validation study of the modelled NaI(Tl) detector has been performed based on the experimental results for absolute detector efficiency values obtained from 226Ra, 232Th, and 40K radioisotopes. Our findings indicate that the obtained absolute detector efficiency values are quite close to used experimental values. Finally, we used the modelled detector for determination of mass attenuation coefficients of Ordinary concrete, Lead, Hematite-serpentine concrete, and Steel-scrap concrete at 186.1, 295.22, 351.93, 609.31, 1120.29, 1764.49, 238.63, 911.2, 2614, and 1460.83 keV gamma-ray energies. Additionally, according to our findings, mass attenuation coefficients obtained from the newly designed detector are compatible with the standard NIST (XCOM) data. To conclude, continuous optimisation procedures are strongly suggested for sophisticated Monte Carlo simulations in order to maintain a high degree of simulation reliability. As a result, it can be concluded that the validation of the simulation model is necessary using measured data. Finally, it can also be concluded that the validated detector models are effective instruments for obtaining basic gamma-ray shielding parameters such as mass attenuation coefficients. © 2022 Huseyin Ozan Tekin et al., published by De Gruyter.
Keywords: ABSOLUTE DETECTOR EFFICIENCY
MCNPX
NAI(TL) DETECTOR
SCINTILLATION DETECTORS
URI: http://elar.urfu.ru/handle/10995/117966
Access: info:eu-repo/semantics/openAccess
SCOPUS ID: 85133564061
WOS ID: 000817962500001
PURE ID: 30621465
ISSN: 23915420
DOI: 10.1515/chem-2022-0169