dc.date.issued2024-
dc.identifier.citationKalyakin A. Stability and Reproducibility of the Amperometric Sensors for Oxygen Concentration Analysis in the Nitrogen Gas Mixtures / A. Kalyakin, A. Volkov, D. Mirzayants // Electrochemical Materials and Technologies. — 2024. — Vol. 3. № 1. — № 20243027.en
dc.identifier.issn2949-0561online
dc.identifier.urihttp://elar.urfu.ru/handle/10995/131610-
dc.descriptionReceived: 26 December 2023. Accepted: 4 March 2024. Published online: 12 March 2024.en
dc.description.abstractThe paper presents experimentally obtained results on the long-term stability tests of the solid electrolyte amperometric sensor output signal during the operation in the O2 + N2 gas mixtures. These data prove the stability and reproducibility of the output signal when measuring the oxygen concentration in air for 8000 hours. The output signal variation during the tests did not exceed ± 2 %. We performed four heating / cooling cycles of different durations, which did not influence either sensor integrity or operation characteristics. The structure of the solid electrolyte sensor and the solid electrolyte/electrode interfacial layer remained unchanged during the tests. Dynamic characteristics of the sensor, including the response time, were stable.en
dc.description.sponsorshipThe present research was performed within the budget plan of the Institute of High Temperature Electrochemistry.en
dc.language.isoenen
dc.publisherУральский федеральный университетru
dc.publisherUral Federal Universityen
dc.relation.ispartofElectrochemical Materials and Technologies. 2024. Vol. 3. № 1en
dc.subjectSOLID ELECTROLYTEen
dc.subjectSENSORen
dc.subjectSTABILITYen
dc.subjectLIMITING CURRENTen
dc.subjectRESPONSE TIMEen
dc.titleStability and Reproducibility of the Amperometric Sensors for Oxygen Concentration Analysis in the Nitrogen Gas Mixturesen
dc.typeArticleen
dc.identifier.rsihttps://www.elibrary.ru/item.asp?id=65429133-
dc.identifier.doi10.15826/elmattech.2024.3.027-
local.issue1-
local.volume3-
local.description.order20243027-
local.contributorKalyakin, Anatoly S.en
local.contributorVolkov, Aleksander N.en
local.contributorMirzayants, Daria P.en
local.contributorКалякин, Анатолий Сергеевичru
Располагается в коллекциях:Electrochemical Materials and Technologies

Файлы этого ресурса:
Файл Описание РазмерФормат 
emt_2024_v3_1_02.pdf523,31 kBAdobe PDFПросмотреть/Открыть


Все ресурсы в архиве электронных ресурсов защищены авторским правом, все права сохранены.