Пожалуйста, используйте этот идентификатор, чтобы цитировать или ссылаться на этот ресурс: http://elar.urfu.ru/handle/10995/109138
Полная запись метаданных
Поле DCЗначениеЯзык
dc.contributor.authorГорбунов, Б. Б.ru
dc.contributor.authorВостриков, В. А.ru
dc.contributor.authorНестеренко, И. В.ru
dc.contributor.authorТелышев, Д. В.ru
dc.contributor.authorGorbunov, B. B.en
dc.contributor.authorVostrikov, V. A.en
dc.contributor.authorNesterenko, I. V.en
dc.contributor.authorTelyshev, D. V.en
dc.date.accessioned2022-03-15T08:25:49Z-
dc.date.available2022-03-15T08:25:49Z-
dc.date.issued2021-
dc.identifier.citationСравнение энергетической эффективности дефибрилляционных импульсов на базе гипотезы гарантированной дефибрилляции / Б. Б. Горбунов, В. А. Востриков, И. В. Нестеренко, Д. В. Телышев // Ural Radio Engineering Journal. — 2021. — Vol. 5, No. 4. — P. 353–368.ru
dc.identifier.issn2588-0454print
dc.identifier.issn2588-0462online
dc.identifier.urihttp://elar.urfu.ru/handle/10995/109138-
dc.descriptionБлагодарим АО «ПО «УОМЗ», в лице главного конструктора КБ медицинских изделий Чупова Алексея Александровича, за активное развитие прикладных исследований в области электрической дефибрилляции.ru
dc.descriptionWe are grateful to JSC PA UOMP, represented by the chief designer of the design bureau of medical devices, Aleksey A. Chupov, for the active development of applied research in the field of electrical defibrillation.en
dc.descriptionПоступила: 27.09.2021. Принята в печать: 12.01.2022.ru
dc.descriptionReceived: 27.09.2021. Accepted: 12.01.2022.en
dc.description.abstractЦелью данного исследования является сравнение на основе гипотезы гарантированной дефибрилляции энергетической эффективности биполярного трапецеидального импульса с фиксированной длительностью фронта и среза с основными типами дефибрилляционных импульсов: классической трапецеидальной (truncated exponential) со спадом вершины 50%, прямоугольной и полусинусоидальной. Результаты компьютерного моделирования на базе гипотезы гарантированной дефибрилляции показали, что дефибрилляционные импульсы энергетически эффективны (имеют низкие значения пороговой энергии дефибрилляции) в достаточно узком диапазоне значений длительности фаз, за пределами которого наблюдается быстрый рост пороговой энергии. По энергетической эффективности трапецеидальный импульс с пологими фронтом и срезом очень близок к полусинусоидальному, и при этом он имеет более широкий диапазон энергетически эффективных длительностей. Существенно более высокую минимальную пороговую энергию гарантированной дефибрилляции имеют прямоугольный и классический трапецеидальный со спадом вершины 0,5 импульсы, при этом классический трапецеидальный импульс имеет более равномерную характеристику в области энергетически эффективных длительностей. Из полученных результатов можно предположить, что максимальную длительность фаз дефибрилляционного импульса следует ограничивать значением не более 9 мс. При этом номинальная выделенная энергия на сопротивлении нагрузки 175 Ом должна составлять не менее 140 Дж. Возможность увеличения длительности импульса без значимого падения его энергетической эффективности позволит обеспечить выделение большей энергии у больных с высоким сопротивлением грудной клетки и, соответственно, большую вероятность проведения успешной дефибрилляции. Указанное выше увеличит также вероятность проведения успешной дефибрилляции у больных при ошибках наложения дефибрилляционных электродов или использовании сухих многоразовых дефибрилляционных электродов.ru
dc.description.abstractThe aim of this study is to compare, on the basis of the guaranteed defibrillation hypothesis, the energy efficiency of a trapezoidal defibrillation pulse with fixed rise and fall times with the main types of defibrillation pulses: truncated exponential with the tilt of 50%, rectangular and half-sine. The study was carried out using the ten Tusscher–Panfilov 2006 human ventricular myocyte model subjected to simulated fibrillation in the BeatBox simulation environment. Depolarizing excitation stimuli with a high frequency were used to simulate fibrillation. The results of computer simulation based on the hypothesis of the guaranteed defibrillation showed that defibrillation pulses are energetically efficient (have low values of threshold energy of defibrillation) in a rather narrow range of phase duration values, beyond which a rapid increase in the threshold energy is observed. In terms of energy efficiency, the trapezoidal pulse with the sloping rise and fall is very close to the half-sine one, and at the same time it has a wider range of energetically effective durations. Significantly higher minimum threshold energy of guaranteed efibrillation is a characteristic of rectangular and truncated exponential pulses, while the truncated exponential pulse has a more uniform characteristic in the area of energetically effective durations. From the results obtained, it can be assumed that the maximum duration of the phases of the defibrillation pulse should be limited to the value of no more than 9ms. In this case, the nominal delivered energy at the load impedance of 175Ω should be at least 140J. The possibility of increasing the pulse duration without a significant drop in its energy efficiency will ensure the delivery of more energy in patients with high transthoracic impedance and, accordingly, a greater probability of successful defibrillation. The above will also increase the probability of successful defibrillation in patients with defibrillation electrodes placement errors.en
dc.description.sponsorshipНастоящие исследования выполнены при реализации проекта «Создание высокотехнологичного производства медицинских изделий для восстановления функции сердца в обеспечение общедоступной дефибрилляции».ru
dc.description.sponsorshipThis research was carried out during the implementation of the project “Creation of high-tech production of medical devices for the restoration of heart function to ensure public defibrillation.”en
dc.language.isoruen
dc.publisherУральский федеральный университетru
dc.publisherUral Federal Universityen
dc.relation.ispartofUral Radio Engineering Journal. 2021. Vol. 5. № 4en
dc.subjectДЕФИБРИЛЛЯЦИЯru
dc.subjectФОРМА ДЕФИБРИЛЛЯЦИОННОГО ИМПУЛЬСАru
dc.subjectЭНЕРГЕТИЧЕСКАЯ ЭФФЕКТИВНОСТЬru
dc.subjectМОДЕЛИРОВАНИЕ, МОДЕЛЬ КАРДИОМИОЦИТАru
dc.subjectЦИКЛ ФИБРИЛЛЯЦИИru
dc.subjectГИПОТЕЗА ГАРАНТИРОВАННОЙ ДЕФИБРИЛЛЯЦИИru
dc.subjectDEFIBRILLATIONen
dc.subjectDEFIBRILLATION PULSE WAVEFORMen
dc.subjectENERGY EFFICIENCYen
dc.subjectSIMULATIONen
dc.subjectCARDIOMYOCYTE MODELen
dc.subjectFIBRILLATION CYCLEen
dc.subjectGUARANTEED DEFIBRILLATION HYPOTHESISen
dc.titleСравнение энергетической эффективности дефибрилляционных импульсов на базе гипотезы гарантированной дефибрилляцииru
dc.title.alternativeComparison of the Energy Efficiency of Defibrillation Pulses Based on the Hypothesis of Guaranteed Defibrillationen
dc.typeArticleen
dc.typeinfo:eu-repo/semantics/articleen
dc.typeinfo:eu-repo/semantics/publishedVersionen
dc.identifier.rsihttps://www.elibrary.ru/item.asp?id=48087952-
dc.identifier.doi10.15826/urej.2021.5.4.002-
local.description.firstpage353-
local.description.lastpage368-
local.issue4-
local.volume5-
local.contributorГорбунов, Борис Борисовичru
local.contributorВостриков, Вячеслав Александровичru
local.contributorНестеренко, Игорь Валерьевичru
local.contributorТелышев, Дмитрий Викторовичru
Располагается в коллекциях:Ural Radio Engineering Journal

Файлы этого ресурса:
Файл Описание РазмерФормат 
urej-2021-4-02.pdf274,04 kBAdobe PDFПросмотреть/Открыть


Все ресурсы в архиве электронных ресурсов защищены авторским правом, все права сохранены.